Bottle Top Dispenser Level 1 Service

Available "ON-SITE" as well as by Mail-in. The choice is yours.

As part of our Level 1 Bottle Top Dispenser Service:

Parts are priced competitively at or below retail price and are subject to change.

Calibrate, Inc. will abide by any and all safety requirements of your company.

Calibrate, Inc. guarantees messages will be answered with a return call from a Calibrate, Inc. Service Representative within 48 hours of an inquiry.

Warranty Declaration for calibration, labor, and parts.


1. Bottle Top Dispenser Identification:

All Bottle Top Dispensers are serviced are identified by serial number, manufacturer, and size. The next date for recommended service will be based on a semi-annual maintenance schedule unless otherwise specified by the client.

2. Preventative Maintenance and Accuracy Verification for Non-GMP/GLP Applications:

  • Complete disassembly (as applicable), inspection, and cleaning of instrument.
  • Replacement of Teflon seal and o-ring or grease, if the instrument requires a seal, and as recommended by the manufacturer.
  • Calibration of instrument to Calibrate, Inc.'s SOP 225–Level 1 specifications.
  • Sticker affixed to Bottle Top Dispenser with date of service and service provider.
  • Toll-Free Service Center telephone number and easy-to-use Service Ordering via our website.
  • All Measurement and Test Equipment used for calibration is SI through NIST traceable (copies of all NIST traceable Calibration Certificates are available upon request).
  • Accuracy verification consists of three readings at three different volumes covering the range of the instrument.


    3. Measurement and Test Equipment Verification and Traceability:

    All Measurement and Test Equipment is calibrated by accredited third party vendors of calibration, and copies of NIST (SI) traceable Calibration Certificates are available upon request.
  • The balance is allowed to equilibrate to the customer's location, and then its performance is verified prior to use. (Reference Calibrate, Inc.'s SOP 181.)
  • Only Class I NIST (SI) traceable weights and NIST (SI) traceable Delta Range balances are used.

    Order Bottle Top Dispenser Service